Major Features
Small in volume、light in weight and convenient in carrying
Operation of the product is simple ,low power consumption
Operation menu for the function setting
SpO2 value display
Pulse rate value display, bar graph display
Pulse waveform display
With backlight
A pulse rate sound indication
With measured data overruns limits and low-voltage alarm function,
the upper/down alarm range is adjustable
Battery capacity indication
Low-voltage indication: low-voltage indicator appears before
working abnormally which is due to low-voltage ,and with alarm
function
Real-time data can be transmitted to computers(Option)
Connected with an external oximeter probe Main performance
Display Mode: LCD display
Screen Resolution: 160*96
SpO2 Measuring Range: 0%~100%, (the resolution is 1%).
Accuracy: 70%~100%2% ,Below 70% unspecified.
PR Measuring Range: 30bpm~250bpm, (the resolution is 1bpm)
Accuracy2bpm or ±2% (select larger)
Measurement Performance in Weak Filling Condition:
SpO2 and pulse rate can be shown correctly when pulse-filling ratio is
0.4%. SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).
Resistance to Surrounding Light: The deviation between the value
measured in the condition of man-made light or indoor natural light
and that of darkroom is less than ±1%.
Power Consumption: less than 80mA
Voltage: DC 3.0V
Power Supply: 1.5V (AA size) alkaline batteries2
Battery Working Hour: Two batteries can work continually for 28
hours(without backlight)
Safety Type: Interior Battery, BF TypeAccessories
Sell in standard:
A user manual
An oximeter probe
Sell in addition:
A data line
A disk (PC software)
Other oximeter probe(Refer to probe application instruction for
details and notice renewal) Physical Identity
Dimension: 109(L)95(W)40(H) mm
Weight: About 215g (with the batteries) Qualification
Passed CE Optional color:Cms60A handheld Pulse oximeter
Technical Parameters
Display Format: LCD;Back light: LED
SPO2:35-99% ; Pulse Ratio:30-250 BPM
Pulse Ratio:30-250 BPM; Spo2 Alarm Range: 0-100%
Pulse Alarm Range: 15-300 BPM
Pulse Wave: LCD Graphic Display
Battery Voltage Low Indication: LCD signals.
Power: CMS60A:Two AA size 1.5V Alkali Battery; CMS60B:3.7V Li-ion Battery
Power Consumption: less than 20 mA.
Resolution: ±1 for SPO2 and ±1 BPM for Pulse Ratio.
Measurement Accuracy: ±2 in stage of 70%-99% SPO2 and meaningless when stage
being smaller than 70%. ±2BPMfor Pulse Ratio.
Measurement Performance in Weak Filling ConditionOxyhemoglobin and
pulse &
#118alue can be shown correctly when pulse-filling ratio is 0.4%.
Interference Resistance Capacity against Ambient LightDeviation is
smaller than ±
1% between #118alue of Oxyhemoglobin measured in natural lighting indoor
condition and present
lighting sources and that measured in dark room.
Option: ECG waveform Display, Heart rate, HR Alarm, SD Card for 24
hours storage,